Model Number 83779
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 22nov2022.It was reported that the coil could not advanced.The target lesion was located in the mildly tortuous and mildly calcified spermatic vein.After a 0.021 microcatheter was placed, a 5mm x 15cm pe f-idc 2d coil was delivered in the microcatheter but the coil got stuck in the opening and could not be pushed out.The coil was pulled back with the microcatheter.The procedure was completed with another of the same device.No complications were reported and the patient's status was stable.However, returned device analysis revealed the interlocking arm was found detached.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.Visual inspection, the main coil, the pusher wire, the introducer sheath were returned.The main coil and the pusher wire were interlocked inside the introducer sheath.The introducer sheath was found with blood residues.Twist lock was opened.The pusher wire was kinked.The main coil was stretched.The interlocking arm was found detached.Functional inspection, the pusher wire was advanced through the introducer sheath, but it was not possible to continue advancing it because the coil was stuck, it was necessary to cut the introducer sheath in order to release the main coil.Microscope inspection, under magnification was possible observed that interlocking arm detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15976160
Report Number2124215-2022-51895
Device Sequence Number5896172
Product Code KRD
UDI-Device Identifier08714729764984
UDI-Public08714729764984
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2024
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0028636894
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2022
Initial Report FDA Received Date12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
Patient Weight72 KG
-
-