| Model Number |
H700977 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, the claris cpu experienced a hard shut down towards the end of the procedure, before performing final testing.The system was able to fully power on and restore case records by unplugging the power from the main cpu as well as from the battery pack/surge protector that it is plugged into.There was no apparent physical contact made to initiate the shutdown.The remainder of the procedure was cancelled, leaving the last part of the ep study (pace testing) unfinished.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h2, h3.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported unexpected shutdown and subsequent cancellation could not be determined.
|
| |
|