MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED) Back to Search Results
Model Number KD-VC631Q-07301S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus that while using the single use preloaded sphincterotome v (distal wireguided), the cutting wire broke.The issue occurred during the endoscopic retrograde cholangiopancreatography (ercp) procedure, there was no delay, and the therapeutic procedure was completed with a similar device.There was no patient harm associated with the event.
 
Additional Manufacturer Narrative
The device was reportedly discarded and is not expected to be returned for evaluation.The subject device was manufactured in july 2022 (within 2 years of reported event) based on the provided 3-digit lot information.However, the manufacture date could not be identified since the subject device was not returned.The device history record for the lot indicated no anomaly with the event-related items.The record includes the following.Process inspection sheet.Quality inspection sheet.Nonconforming product report.The event can be detected/prevented by following the instructions for use which state: before use, prepare and inspect the instrument and a cord as instructed below.Should any irregularity be observed, do not use the instrument or a cord; use a spare instead.Damage or irregularity may compromise patient or user safety, pose an infection control risk, cause tissue irritation, perforations, bleeding, mucous membrane damage, thermal injury, and may result in more severe equipment damage.Do not use excessive force to bend the distal end of the sphincterotome.This could damage the cutting wire.Do not operate the sphincterotome if the insertion portion is bent or the distal end of the tube is straight.This could damage the tube.Do not move the slider rapidly.This could damage the sphincterotome.Do not stretch the cutting wire too tightly.This could damage the instrument.Since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.·do not insert the instrument into the endoscope when the cutting wire is tightened.Doing so may extend the distal end of the insertion portion from the endoscope tip abruptly.This could cause patient injury, such as perforations, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.Do not force the instrument if resistance to insertion is encountered.Reduce the angulation or lower the forceps elevator of the endoscope until the instrument passes smoothly.The use of excessive force could cause patient injury, such as perforations, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.Do not tighten the cutting wire more than necessary, and do not force it against the papilla of vater.This could result in patient injury, such as perforations, bleeding, or mucous membrane damage.Do not withdraw the instrument from the endoscope while the cutting wire is tightened.This could damage the endoscope or instrument.The exact cause of the reported phenomenon could not be conclusively identified by the following reason: ·the device was not returned for the investigation.·the device history record for the lot indicated no anomaly with the event-related items.Based on the investigation result of the similar complaint in the past, it can be inferred that an excessive force was applied to the cutting wire after it was bent near the distal tip.This might have caused the cutting wire to break and fall off.However, the circumstances surrounding the event were unclear, and the exact cause could not be determined.Although the root cause of this event could not be determined, olympus will continue to monitor field performance for this device, and take appropriate actions as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED)
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15977016
Report Number9614641-2022-00765
Device Sequence Number20556572
Product Code KNS
UDI-Device Identifier04953170371493
UDI-Public04953170371493
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141991
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-VC631Q-07301S
Device Lot Number27V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Initial Report FDA Received Date12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-