AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-XP150N
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device referenced in this report was not returned for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
The olympus representative reported that during a diagnostic upper endoscopy procedure, the gastrointestinal videoscope presented with a lateral noise that entered the octagonal screen of the endoscope during the procedure and displayed error code e311 (white balance not set).The endoscope screen had appeared and disappeared.There was a possibility that the contact on the scope cable was poor, so the user tried disconnecting and reconnecting the cable, but the issue did not improve.Therefore, when checked using a simple water leakage checker, it was found that air and water leakage bubbles were coming out of the scope connector stopper, and there was a gap.The procedure was cancelled.The customer confirmed there was no health hazard to the patient.
 
Event or Problem Description
Additional information was received from the customer.The procedure was for a screening test with no health hazard for the patient.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information from the customer.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 7 years since the subject device was manufactured.Based on the results of the investigation, water intrusion inside the scope likely caused moisture to attach to the circuit board of the electrical circuit board.As a result, a conduction abnormality such as a short circuit occurred.In addition, external stress likely caused the connecting section of the parts to break affecting the weather tightness of the device.However, the root cause of the reported event could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15977277
Report Number9610595-2022-05423
Device Sequence Number6894983
Product Code FDS
UDI-Device Identifier04953170307478
UDI-Public04953170307478
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051645
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-XP150N
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer12/15/2022
01/23/2023
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date12/16/2022
02/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/07/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-