| Model Number |
01.29.206 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 21-nov-2022: lot 1907349: (b)(4) items manufactured and released on 08-jan-2020.Expiration date: 2024-12-11 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.Other device involved: batch review performed on 21-nov-2022.Liner: cc e cc light 01.26.3244hct flat pe hc liner ø 32 / e (k103352) lot 2009637: (b)(4) items manufactured and released on 17-nov-2020.Expiration date: 2025-11-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.
|
| |
|
Event or Problem Description
|
|
Revision surgery performed due to joint luxation 1 year and 8 months after the primary surgery.The patient did an hormone therapy to get pregnant which results in a slackening of the tissue.
|
| |
|