BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Catalog Number 114362 |
Device Problem
Inadequate Ultra Filtration (1656)
|
Patient Problems
Chest Pain (1776); Red Eye(s) (2038); Swelling/ Edema (4577)
|
Event Date 11/21/2022 |
Event Type
Injury
|
Event Description
|
It was reported that two hours into treatment with an ak 96 machine, the patient presented with red and swollen eyes, congested eyeballs, swollen stomach, and chest tightness.The treatment was ended immediately.The post treatment weight was reported to be 57.5 kg indicating a 5.9 kg weight gain.Treatment was restarted with the different dialysis machine to remove excess fluid.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information added to h3, h6 and h10.The device was evaluated onsite by a local technician.The machine was found working as per specifications.The reported condition was not duplicated, no information was provided suggesting any defect or malfunction of the used machine.The reported condition was not verified.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|