• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114362
Device Problem Inadequate Ultra Filtration (1656)
Patient Problems Chest Pain (1776); Red Eye(s) (2038); Swelling/ Edema (4577)
Event Date 11/21/2022
Event Type  Injury  
Event Description
It was reported that two hours into treatment with an ak 96 machine, the patient presented with red and swollen eyes, congested eyeballs, swollen stomach, and chest tightness.The treatment was ended immediately.The post treatment weight was reported to be 57.5 kg indicating a 5.9 kg weight gain.Treatment was restarted with the different dialysis machine to remove excess fluid.No additional information is available.
 
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information added to h3, h6 and h10.The device was evaluated onsite by a local technician.The machine was found working as per specifications.The reported condition was not duplicated, no information was provided suggesting any defect or malfunction of the used machine.The reported condition was not verified.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15978461
MDR Text Key305457320
Report Number9616026-2022-00114
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number114362
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-