| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
03/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-03473; 0001822565-2022-03474; 0001822565-2022-03475; 0001822565-2022-03476; 3007963827-2022-00325; 0001822565-2022-03478; 0001822565-2022-03479; 0001822565-2022-03481; 0001822565-2022-03483; 0001822565-2022-03484.
|
| |
|
Event or Problem Description
|
|
It was reported a patient underwent an initial left knee arthroplasty with unknown product.The patient was revised approximately 1.5 years later due to implant loosening.At the 1 year follow up, the patient presented with left lateral knee pain that radiates to lower leg.Narcotics were prescribed.At the 2 year follow up appointment.A large left knee effusion was present.Left knee aspiration was completed in the office.No further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a total knee arthroplasty.At the 1 year follow up, the patient presented with left lateral knee pain that radiates to lower leg.Narcotics were prescribed.At the 2 year follow up appointment, a large left knee effusion, tibial radiolucency, and laxity was present.Left knee aspiration was completed in the office.No further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified moderate pain with some effusion but no instability.X-ray found tibial radiolucency.A definitive root cause cannot be determined.Per package insert: pain, swelling are known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receiving additional information of the reported event, it was determined to be not reportable.The initial report should be voided as only the articular surface was revised.
|
| |
|
Event or Problem Description
|
|
Upon receiving additional information of the reported event, it was determined to be not reportable.The initial report should be voided as only the articular surface was revised.
|
| |
|
Search Alerts/Recalls
|