| Model Number |
24690 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 3.0mmx20mmx135cm (4f) sterling balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.Another 3.0mmx40mmx135cm (4f) sterling balloon catheter was advanced, but the balloon also ruptured during inflation.Both devices were removed with no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 3.0mmx20mmx135cm (4f) sterling balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.Another 3.0mmx40mmx135cm (4f) sterling balloon catheter was advanced, but the balloon also ruptured during inflation.Both devices were removed with no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a longitudinal tear in the balloon 7mm from the tip and is approximately 40mm long.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a longitudinal tear in the balloon.
|
| |
|