Model Number 24690
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.A 3.0mmx20mmx135cm (4f) sterling balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.Another 3.0mmx40mmx135cm (4f) sterling balloon catheter was advanced, but the balloon also ruptured during inflation.Both devices were removed, and no portions were left behind.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a longitudinal tear in the balloon 6mm from the tip and is approximately 21mm long.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a longitudinal tear in the balloon.
 
Event or Problem Description
It was reported that balloon rupture occurred.A 3.0mmx20mmx135cm (4f) sterling balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.Another 3.0mmx40mmx135cm (4f) sterling balloon catheter was advanced, but the balloon also ruptured during inflation.Both devices were removed, and no portions were left behind.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key15978677
Report Number2124215-2022-50076
Device Sequence Number17873213
Product Code DQY
UDI-Device Identifier08714729755586
UDI-Public08714729755586
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K141150
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2024
Device Model Number24690
Device Catalogue Number24690
Device Lot Number0028365998
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date04/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INFLATION DEVICE: ENCORE.; INFLATION DEVICE: ENCORE.
Patient SexUnknown
-
-