MAUDE Adverse Event Report: ENCORE MEDICAL L.P. GUIDE; GUIDE, REAMER, BONE TAP, 6.5MM, RSP, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. GUIDE; GUIDE, REAMER, BONE TAP, 6.5MM, RSP, SHOULDER Back to Search Results
Model Number 804-03-008
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
Instrument failure: the tap broke in half inside of the brown reamer.Part of the broken tap is stuck inside of the brown reamer, leaving both instruments broken.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
The reason for this instrument failure was reported as broken tap.The healthcare professional indicated this event occurred during surgery, near the patient.No response received by the surgeon.The surgery was completed as intended, with a twenty-minute delay.The instrument was inspected prior to use and was deemed acceptable for use based on its appearance.The agent was present during surgery and was able to source a suitable replacement device.The devices were returned to the manufacturer and made available at djo surgical for evaluation.The revision level or lot number were not reported; therefore, this instrument could not be linked to a specific device history record (dhr) or the actual date of manufacture could not be determined with confidence.Complaint database review showed 31 previous complaints but there were no indications that this instrument has a design or material deficiency.Those are 29 - broke/cracked/damaged, 1 - device cracked/ broke and 1 - threads damaged/galled.The root cause of this complaint is likely attributable to damage incurred from prolonged use and through misuse or rough handling which surgical instruments are subjected to.This is not an event associated with a product failure, malfunction, or issue.There are no indications that this instrument has a systemic design or material deficiency.Therefore, no containment of inventory is required.Event is associated with instrument usage, not a design or manufacturing issue.Rma examination: the instrument was returned to djo and after further examination, tap broke in half.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE
Common Device Name
GUIDE, REAMER, BONE TAP, 6.5MM, RSP, SHOULDER
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15980335
Report Number1644408-2022-01674
Device Sequence Number14923609
Product Code FZX
UDI-Device Identifier00888912147842
UDI-Public(01)00888912147842
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number804-03-008
Device Catalogue Number804-03-008
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer01/06/2023
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date01/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
804-06-012 LOT 256639L01.; 804-06-012 LOT UNKNOWN.
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-