| Model Number |
SSM-640 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.The event of "capsular contracture, baker grade iii/iv" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device rupture and capsular contracture, baker grade iii/iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "unilateral left-side capsular contraction, baker grade iii/iv implant exchange." healthcare professional additionally reported "left side rupture confirmed with mri." device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, three openings in posterior side assessed as surgical damage.Capsular contracture: unable to observe as it is a medical event not related to the device.Additional observations: creases is observed.White particles were observed on the shell.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|