• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO FATHOM SYSTEM; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO FATHOM SYSTEM; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 29462
Medical Device Problem Codes Difficult to Advance (2920); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 29nov2022.It was reported that the microcatheter could not cross the imaging catheter and was kinked.The target lesion was located in the liver.A 180cm renegade hi-flo 16 was selected for use.During the procedure, it was noted that the microcatheter set could not cross the imaging catheter.After several attempts, the device got kinked, therefore, it could not be used.The procedure was completed with another of the same device.There were no complications reported and the patient was stable post procedure.However, device analysis revealed that the device fractured 40.5cm from the hub.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The hub and shaft were visually inspected when returned.The renegade device showed multiple small bends and kinks.The device showed stretching and a fracture located 40.5cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The advancing issues could not be confirmed due to the damage on the catheter shaft.The complaint was confirmed for shaft damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE HI-FLO FATHOM SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15981141
Report Number2124215-2022-51209
Device Sequence Number12236280
Product Code KRA
UDI-Device Identifier08714729790853
UDI-Public08714729790853
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140329
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2025
Device Model Number29462
Device Catalogue Number29462
Device Lot Number0029437873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/29/2022
Initial Report FDA Received Date12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale