Model Number TN FC500 FLOW CYTOMETER 100-240V 50/60HZ
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
The field service engineer (fse) reported signal shift at fl2 and fl3 positions on the fc500 flow cytometer while running flow check control while servicing the instrument.There was no report of death, injury, or change to patient treatment as a result of this event.There were no erroneous patient results generated.
 
Additional Manufacturer Narrative
Patient information is not applicable.There was no impact to patients as a result of this event.While the fse was servicing the instrument for an unrelated issue, he observed that the tarpon xl amp assembly at the fluorescence channel (fl2 and fl3) positions was defective.To resolve the issue the fse replaced the board at the identified positions.Bec is filing an mdr for this event based on the fda classification of the 08 jan 2018 urgent medical device recall as a class i recall on 20 nov 2018 (recall number z-0471-2019 for fc 500; recall number z-0472-2019 for epics xl/xlmcl).Bec internal identifier: (b)(4).
 
Additional Manufacturer Narrative
Follow up report 01- correction field g3 date received by manufacturer revised to reflect correct date of awareness by bec.Bec internal identified - (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FC 500 FLOW CYTOMETER
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
angela vettel
250 s. kraemer blvd
brea, CA 92821
7149613625
MDR Report Key15982324
Report Number1061932-2022-00075
Device Sequence Number6960534
Product Code GKZ
UDI-Device Identifier15099590572839
UDI-Public(01)15099590572839(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K071681
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall,Modification/Adjustment,Replace,Notification
Type of Report Initial,Followup
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTN FC500 FLOW CYTOMETER 100-240V 50/60HZ
Device Catalogue Number626553
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date12/14/2022
Supplement Report FDA Received Date02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2050012-0108/2018-001C
Patient Sequence Number1
Patient SexUnknown
-
-