Model Number 26280
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the stent failed to deploy.The patient presented for lower extremity venous balloon dilation and stenting of the target lesion located in the inferior vena cava.A16x90/9fr uni plus 75cm wallstent endoprosthesis stent as advanced for treatment but could not deploy.The procedure was completed with another of same device.No patient complications were reported and the patient status was stable.However, returned device analysis revealed a stent damage.
 
Additional Manufacturer Narrative
Device evaluation by mfr: a wallstent uni was returned for analysis.The stent was returned partially deployed by approximately 10mm.The stent was noted to be damaged at the distal end.The investigator was unable to deploy the stent due to the kink in the shaft.A visual examination identified no damage or issues with the stent cups or stent holder of the returned device.A visual and tactile examination identified no issues with the tip of the device.A visual and tactile examination identified a severe shaft kink 50mm distal from the distal end of the strain relief.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15983245
Report Number2134265-2022-08958
Device Sequence Number5884019
Product Code MAF
UDI-Device Identifier08714729204053
UDI-Public08714729204053
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2023
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0028245453
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2022
Initial Report FDA Received Date12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
Patient Weight75 KG
-
-