We checked the returned unit and confirmed that the nozzle gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the nozzle gluing.In addition, we confirmed that the light guide fiber bundle (lcb) broken, the remote control buttons cracked, the lg cable connector corroded, the lg air/water socket cylinder corroded, the operation channel connector dirty, and the bending rubber gluing poor finish; however, they are not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, no response was obtained, so we cannot deny the possibility of the glue falls into the human body.Therefore, based on the technical report "hr-rpt-0585(nozzle)" and/or the risk analysis results, it was evaluated to submit mdr.
|