Manufacturer's ref#:(b)(4).Technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.No investigation done on actual device.Trended case results: the charger has been returned, because the micro usb interface has broken down, causing the device to become hot.The charger and cable shows visible signs of heating.The area around the connector is deformed, as well as the cable connector.A substance is seen inside the connector and also below the connector on backside housing.The components on pcb around the connector seems fine.Functional testing reveals that the charger works fine, when running on the internal battery.That means that the malfunction in the device is purly a breakdown of the micro-usb connector.The connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro-usb is a very well known and used standard but can mechanically break down.The clinical investigation concluded: the connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.Based on these results from previous evaluations, clinical conclusion on this is that there is a potential risk of heat dissipation based on the findings.A damaged power supply or a power supply of very poor quality could lead to burn injuries and in rare conditions there is a small risk that a wrong charging or discharging or a battery thermal runway can lead to high temperatures or fire accidents.The chargers have been safety tested according to applicable standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.There are no new clinical aspects (e.G.Unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusion herein are sufficient.Risk register reference:the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.Manufacturer's investigaton is final.
|