| Model Number |
RA22-2BB BENT FLANGE TIP |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information were not provided.Sorin group italia manufactures the pureflex® arterial cannula ra22-2bb bent flange tip.The incident occurred in poland.The involved cannula is not available for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Sorin group italia has received a report that, at the end of a procedure, the body of the cannula was severely damaged.There is no report of any patient injury.According to the surgeon they used the same clamps as usual.
|
| |
|
Event or Problem Description
|
|
See intial report.
|
| |
|
Additional Manufacturer Narrative
|
|
The dhr review highlighted that the lot whose claimed product belong to was released as conform according to specifications.An exhaustive photo was provided as evidence of the defect, showing a failure of the cannula body material near the clamping area.The complaints' review performed was able to identify no other similar issue on the product part number ra2x-xxx in the 3 years.Thus, the event is isolated.Considering that the lot involved was (b)(4) units distributed worldwide, and no other similar event was recorded either on the affected lot or on the affected code in the last three years, a manufacturing failure of the cannula can be resonably ruled out.However, considering that the defect occurred at the end of the procedure, it cannot be excluded that the material deteriorated progressively during use most likely because of inadvertidely initial rough handling of user.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|
Search Alerts/Recalls
|