It was reported that, after a tka surgery performed using the lgn ps high flex xlpe sz 5-6 9mm device, the patient presented medial joint pain/laxity.Due to this, a revision surgery was performed to exchange the lgn ps high flex xlpe sz 5-6 9mm device to a constrained insert.Further details not available yet.
|
|
Section h3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided for evaluation.Therefore, there were no clinical factors found which would have contributed to the reported event.The patient's current condition is unknown and the patient impact beyond the revision surgery could not be determined.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for knee system revealed that looseness of components has been identified in possible adverse effects section as a result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions.It also states that temporary or permanent nerve damage can result in pain or numbness of the affected limb.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.Factors that could contribute to the reported event include traumatic injury, joint tightness or patient condition.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|