| Catalog Number |
328520 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the relion insulin syringes 1/2ml 31 gauge, 6mm (15/64") went through the shield after recapping.The following information was provided by the initial reporter: consumer reported that the needle of one syringe went through the shield and stuck her finger as she was trying to re-shield after injection.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the device history record was completed for batch# 2178998.All inspections and challenges were performed per the applicable operations qc specification.
|
| |
|
Event or Problem Description
|
|
It was reported that the relion insulin syringes 1/2ml 31 gauge, 6mm (15/64") went through the shield after recapping.The following information was provided by the initial reporter: consumer reported that the needle of one syringe went through the shield and stuck her finger as she was trying to re-shield after injection.
|
| |
|