• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE 132 SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE 132 SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6020-4535
Medical Device Problem Codes Degraded (1153); Material Erosion (1214)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6) 2016 and was revised on (b)(6) 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
Reported event: an event regarding wear/metallosis and abnormal ion level involving an accolade stem that was mated with an lfit v40 cocr head was reported.The wear event was confirmed via evaluation of the provided photograph of the stem and clinician review of the provided medical records.Abnormal ion level was not confirmed.Method & results: product evaluation and results: the reported device was not returned however a photograph of the stem was provided for review.Visual inspection of the provided photograph indicated that the stem is severely worn consistent with loss of taper lock.Clinician review: a review of the provided medical records by a clinical consultant indicated: "this patient underwent a primary total hip arthroplasty in 2009 and then approximately 13 years later developed elevated metal ions, metallosis and trunnion failure requiring revision.I can confirm that the patient underwent the primary and revision procedure since i was able to review the operation reports.The root cause of this event cannot be determined with certainty.Causes of elevated ion levels, trunnion failure and metallosis are multifactorial including surgical technique, patient activity level and bmi, as well as implant factors." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to elevated metal ion levels and severe wear of the stem trunnion at the stem-head junction.The reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about on (b)(6) 2016 and was revised on (b)(6) 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in his blood.Update per medical records, 9/22/2023: event is for the right hip, not left."gross metal debris encountered in the hip, with black metal staining of the periprosthetic tissues.The musculature around the hip was affected including portions of the abductors.Then the prior femoral head was removed with a bone tamp.The trunnion had grossly failed and deformed and was not salvageable.A synovectomy and debridement of the periprosthetic tissues and hip joint was performed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE 132 SIZE 4.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key15985999
Report Number0002249697-2022-01832
Device Sequence Number5995863
Product Code LPH
UDI-Device Identifier04546540510631
UDI-Public04546540510631
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Catalogue Number6020-4535
Device Lot Number26302703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer10/03/2023
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date10/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/24/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
-
-