Model Number 20690902
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022 charger caught on fire while charging when patient was charging aid in patients room.The patient was not harmed.The hearing aid did not work afterwards.Both hearing aid and chargers were replaced.Original accessories/equipment was used.
 
Additional Manufacturer Narrative
Manufacturer's ref# (b)(4).Technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.Trended case investigation concluded: the connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.The clinical investigation concluded: a damaged power supply or a power supply of very poor quality could potentially lead to burn injuries and in rare conditions there is a small risk that a wrong charging or discharging or a battery thermal runaway can lead to high temperatures or fire accidents.The chargers have been safety tested according to applicable standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.There are no new clinical aspects (e.G.Unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusion herein are sufficient.Risk register reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.This is a final report.No further follow up is expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELTONE
Common Device Name
HEARING AID, AIR-CONDUCTION WITH WIRELESS TECHNOLOGY
Manufacturer (Section D)
GN HEARING A/S
lautrupbjerg 7
ballerup, 2750
DA  2750
Manufacturer Contact
karen frisk
lautrupbjerg 7
ballerup, 2750
DA   2750
MDR Report Key15986395
Report Number3005650109-2022-00081
Device Sequence Number5950182
Product Code OSM
UDI-Device Identifier05708296145638
UDI-Public05708296145638
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number20690902
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Initial Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-