• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; CARDIAC COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; CARDIAC COMPRESSOR Back to Search Results
Model Number RMU-1000
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 07/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the device's electronic history file showed the device was manufactured in december 2020 and during its service, had performed 32,059 compressions.Analysis of the returned device identified the cause of the reported problem was caused by the loosening of the compression module's bottom screw.
 
Event or Problem Description
An international distributor relayed a report from a professional user that the compression device stopped performing compressions during use.It was also reported that there was no death or serious injury associated with the reported device problem.No other patient or event details were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
CARDIAC COMPRESSOR
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
741 boston post road
suite 201
guilford CT 06437
Manufacturer Contact
ian white
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key15986500
Report Number3003521780-2022-00010
Device Sequence Number6209241
Product Code DRM
UDI-Device Identifier10815098020628
UDI-Public10815098020628
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K141809
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRMU-1000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Initial Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/26/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-