| Model Number |
RMU-1000 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
07/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the device's electronic history file showed the device was manufactured in december 2020 and during its service, had performed 32,059 compressions.Analysis of the returned device identified the cause of the reported problem was caused by the loosening of the compression module's bottom screw.
|
| |
|
Event or Problem Description
|
|
An international distributor relayed a report from a professional user that the compression device stopped performing compressions during use.It was also reported that there was no death or serious injury associated with the reported device problem.No other patient or event details were provided.
|
| |
|