ARTIVION, INC. ¿ AUSTIN ONX ASCENDING AORTIC 23; HEART-VALVE, MECHANICAL Back to Search Results
Model Number ONXAAP-23
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Arrhythmia (1721); Atrial Fibrillation (1729)
Date of Event 03/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This investigation is currently ongoing.Any additional information will be provided in the follow-up report.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to artivion ¿ formerly cryolife/jotec is accurate or has been confirmed by artivion ¿ formerly cryolife/jotec.
 
Event or Problem Description
According to aap study - ascend - post market clinical follow-up study in patients treated with the on-x ascending aortic prosthesis (aap) adverse event report, event details: implant date - (b)(6) 2021.Patient age at time of event: 64 yo.Adverse event classification: this event is not linked to the medical device ¿ non-serious (ae).Detailed description of event: no information provided.Outcome of event: no information provided.Information on taken actions because of event: no information provided.Assessment by investigator: no information provided.Adverse event - cardiac arrhythmia.Classification ¿ adverse event.Start date ¿ ((b)(6), 2021.End date ¿ (b)(6), 2021.Duration of event ¿ 0.Outcome ¿ symptoms resolved without sequelae.Causal relationship to the procedure ¿ possible.Causal relationship to the aortic valve disease of heterogeneous ethology ¿ possible.Causal relationship to on-x aap ¿ possible.Number of reoperations performed: 0.This investigation is relegated to onxaap-23 sn(b)(4).
 
Additional Manufacturer Narrative
H10 - additional manufacturer narrative.The date of implant is (b)(6) 2021, corrected from (b)(6) 2021.
 
Additional Manufacturer Narrative
According to initial reports, patient in ascend - post market clinical follow-up study in patients treated with the on-x ascending aortic prosthesis (aap), experienced cardiac arrhythmia.This investigation is relegated to onxaap-23 sn(b)(6).The product will not be returned to the manufacturer for evaluation as it remains implant.Additional information regarding this event was received.The manufacturing records for the onxaap-23 sn (b)(6) were reviewed and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications.A review of the available information was performed.An onxaap 23 with serial number (b)(6)was implanted on (b)(6) 2021 in a 64-year-old male with a past medical history of tobacco abuse, obesity, hypertension, hyperlipidemia, coronary artery disease and atrial fibrillation with a direct current cardioversion [dccv] on (b)(6) 2021, severe aortic stenosis with bicuspid valve and proximal ascending aortic dilation.On (b)(6) 2021 (5 days post implant) the patient was noted to have a heart rhythm on the cardiac monitor of atrial fibrillation with rvr in the 140¿s and reported palpitations.His metoprolol was increased to 25mg twice daily and he was given an iv amiodarone bolus and started on an oral amiodarone titration regimen.The event lasted less than one day per the medical records.On his follow up visit on (b)(6) 2022 (338 days post implant) the notes state that the patient has no recurrent episodes of atrial fibrillation and is no longer on amiodarone.The instructions for use [ifu] for the aap valve list cardiac arrhythmia as a potential adverse event for mechanical prosthetic valve recipients.There is no indication that the on-x valve contributed in any way to the reported episode and it is more likely a complication related to the surgical procedure itself rather than the valve.Cardiac arrhythmia is a known potential complication, however, there is no indication that the on-x valve contributed in any way to the reported episode and it is more likely a complication related to the surgical procedure itself rather than the valve.It should be noted that the patient has a medical history of atrial fibrillation requiring dccv prior to the implant of the onxaap device.No further action is required.Based on the available information, a definitive root cause for this event cannot be determined.Additionally, without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.The manufacturing records were reviewed, and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications.This event does not identify additional hazards or modify the probability and severity of existing hazards.There is no indication that an error or deficiency occurred at artivion ¿ formerly cryolife/jotec and the ifu adequately communicates risk.This complaint was reviewed for a capa evaluation and a capa is not warranted at this time.Artivion will continue to monitor similar complaints to determine if additional actions are warranted; however, at this time no further actions are necessary.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to artivion ¿ formerly cryolife/jotec is accurate or has been confirmed by artivion ¿ formerly cryolife/jotec.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONX ASCENDING AORTIC 23
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ARTIVION, INC. ¿ AUSTIN
1300 e. anderson ln., bldg. b
austin TX 78752
Manufacturer (Section G)
ARTIVION, INC. ¿ AUSTIN
1300 e. anderson ln., bldg. b
austin TX 78752
Manufacturer Contact
rochelle maney
1655 roberts blvd
kennesaw, GA 30144
7704193355
MDR Report Key15986517
Report Number1649833-2022-00068
Device Sequence Number14951290
Product Code LWQ
UDI-Device Identifier00851788001525
UDI-Public851788001525
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P000037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberONXAAP-23
Device Catalogue NumberONXAAP-23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date11/17/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer11/17/2022
11/17/2022
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date02/06/2023
02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
-
-