• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL 45 CM SHAFT, CURVED JAW; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL 45 CM SHAFT, CURVED JAW; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLX145C
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 10/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that post-op to the sleeve gastrectomy procedure, the team coordinator stated they had an enseal that according to the doctor, it did not seal vessels properly in the case.He had to bring the patient back for bleeding today.It was unknown how the case was completed and if there were any patient consequences.No other details could be provided at this time.
 
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.Additional information was requested and the following was obtained: the device caught a little but it sealed okay.The procedure began with the one nslx145c and then was completed with a second nslz145c.Also reported was that there were no patient consequences.It was shared that the patient was brought back to the or.Which vessel(s) was both the devices used on to seal? short gastrics.What other instruments were used in the original procedure? please be more specific; a signia stapler, endo stitch, etc.Was the end tone heard during each sealing sequence when using the device in the original procedure? unknown.Did the surgeon experience any generator alert screens during the original procedure? unknown.Was the source of the bleeding identified? unknown.If yes, was that source where the enseal was used? were similar bleeding issues that were found post-op occur during the original procedure? unknown.How was the bleeding addressed? unknown.Were any blood transfusions required? unknown.What is the patient's current status? surgeon shared that the patient was doing well.Can the surgeon confirm that the bleeding that required the patient to be brought back to the or was not from the short gastrics? doctor's pa said that the bleeding did come from the short gastrics.An analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL 45 CM SHAFT, CURVED JAW
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15986611
Report Number3005075853-2022-08652
Device Sequence Number5724179
Product Code GEI
UDI-Device Identifier20705036015496
UDI-Public10705036015499
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K172580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNSLX145C
Device Catalogue NumberNSLX145C
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/12/2022
Initial Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-