| Model Number |
M00565040 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Imdrf device code a150207 captures the reportable event of delivery system difficult to remove.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2022 that a wallflex enteral colonic stent was implanted to treat a 2-3cm malignant stricture during a colonoscopy with stent placement procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous and was not dilated prior to stent placement procedure.During the procedure, the stent was positioned correctly for deployment; however, it was not completely released as the delivery system became stuck.The physician moved the delivery system several times and it was eventually removed.The wallflex enteral colonic stent remained implanted and the procedure was completed.There were no patient complications reported as a result of this event.
|
| |
|