BOSTON SCIENTIFIC CORPORATION WALLFLEX COLONIC; STENT, COLONIC, METALIC, EXPANDABLE Back to Search Results
Model Number M00565040
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Imdrf device code a150207 captures the reportable event of delivery system difficult to remove.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a wallflex enteral colonic stent was implanted to treat a 2-3cm malignant stricture during a colonoscopy with stent placement procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous and was not dilated prior to stent placement procedure.During the procedure, the stent was positioned correctly for deployment; however, it was not completely released as the delivery system became stuck.The physician moved the delivery system several times and it was eventually removed.The wallflex enteral colonic stent remained implanted and the procedure was completed.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX COLONIC
Common Device Name
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15987189
Report Number3005099803-2022-07340
Device Sequence Number5950215
Product Code MQR
UDI-Device Identifier08714729456513
UDI-Public08714729456513
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K061877
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2024
Device Model NumberM00565040
Device Catalogue Number6504
Device Lot Number0030382788
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/24/2022
Initial Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-