| Model Number |
9-ASD-008 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, an 8mm amplatzer septal occluder was selected for implantation using a 6f amplatzer torqvue delivery system.During the procedure, during deployment, it was discovered that the left atrial disc was deformed into a cobra shape.The device was removed from patient anatomy, and a 9mm amplatzer septal occluder was successfully implanted.The device was not released inside patient anatomy, there was no angulation or kink noted in the delivery system, and there was no interaction with cardiac structures during deployment.The patient is reported to be discharged.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.An image was received from the field and the image appears to show the device deforming cobra-like shape, however, the device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|