• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 45°, 460 MM, QUICK LOCK, AUTOCLAVABLE; RIGID OPTICAL LAPAROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 45°, 460 MM, QUICK LOCK, AUTOCLAVABLE; RIGID OPTICAL LAPAROSCOPE Back to Search Results
Model Number WA52006A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During inspection and testing, the optical system had sunk down.The reported issue (foggy image) was confirmed due to debris on the optical system.In addition, there were scratches on the outer tube.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported the image from the subject device was foggy when entering the patient during a laparoscopic sleeve procedure.The procedure was completed using another similar device and there was no effect on the patient due to the event.The subject device was sent to an olympus service center for evaluation.During inspection and testing, service found the optical system had sunk down.This report is being submitted for the malfunction found during evaluation (sunk down optical system).
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the broken components could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 45°, 460 MM, QUICK LOCK, AUTOCLAVABLE
Common Device Name
RIGID OPTICAL LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, - 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15987284
Report Number9610773-2022-00658
Device Sequence Number6209270
Product Code GCJ
UDI-Device Identifier04042761052726
UDI-Public04042761052726
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K923982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA52006A
Device Catalogue NumberWA52006A
Device Lot Number652755
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer03/23/2023
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date04/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS 4K CAMERA
Patient SexUnknown
-
-