| Model Number |
RUBELLA IGG |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys rubella igg immunoassay results for 3 patient samples on a cobas e 801 analytical unit compared to a siemens assay.For patient 1, the elecsys rubella igg result was 17.2 iu/ml, interpreted as positive.The siemens assay result was negative.For patient 2, the elecsys rubella igg result was 91.9 iu/ml, interpreted as positive.The siemens assay result was negative.For patient 3, the elecsys rubella igg result was 27.2 iu/ml, interpreted as positive.The siemens assay result was negative.The initial elecsys assay results were reported outside of the laboratory.The e801 analyzer serial number is (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The patient samples were provided for investigation.The investigation obtained reactive results for rubella igg for all three samples.The samples were also tested for rubella igm and all three samples were non-reactive.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration was last performed on (b)(6) 2023.The investigation determined that the customer's results were reproducible and the patient samples provided were correctly positive for rubella igg.The reagent performed within specification.
|
| |
|