| Model Number |
IPN915184 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reported issue: the clips are poorly packed and open.Found prior to use.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).P/n 544240 is not being manufactured currently, however, another part number from the same family was use for the verification of failure mode reported in the current manufacturing process and during the test issue reported incomplete seal - sterility compromised was not observed in the current manufacturing process.The device history review for the product hemolok l clips 6/cart 84/box lot# 73d2200552 investigation did not show issues related to the complaint.The customer complaint cannot be confirmed and corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation, confirm the alleged defect and determine the root cause.At this time due the sample is not available is not possible to determine the source of the defect reported.
|
| |
|