• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL POSTERIOR AUGMENT CEMENTED SIZE 11, 11+ 5 MM THICKNESS; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL POSTERIOR AUGMENT CEMENTED SIZE 11, 11+ 5 MM THICKNESS; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 08/15/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a left knee surgery.Subsequently, the patient had swelling in the left knee resulting in a left knee aspiration approximately 2.5 years later.No further complications or results were reported.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 42504607001 - femur cemented standard left size 11 - 64655857.42560117518 - stem extension straight splined uncemented 18 mm diameter +175 mm length - 64716473.42542007901 - tibia fixed cemented left size g stem extension use required - 64679178.42512801012 - articular surface fixed bearing constrained condylar knee (cck) left 12 mm height use with tibia sizes g, h / revision femur sizes 11, 11+, 13 with lo - 64341097 - 42560317514 - stem extension 3mm offset splined uncemented 14 mm diameter +175 mm length - 64739783.42545000513 - tibial central cone size large - 64800272.Unknown - unknown patella - unknown.Unknown - unknown palacos cement - unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-03501, 0001822565-2022-03503, 0001822565-2022-03507, 0001822565-2022-03509, 0001822565-2022-03510, 3005751028-2022-00022, 0001822565-2022-03511.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the patient works as a structural maintenance journeyman, pain for over 1 year, effusion resulting in aspiration relieves pain, no synovial fluid culture.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was further reported that the patient continued to present with recurrent pain and swelling in the knee as well as instability and laxity.The patient was then revised and during the revision, the femoral component was found grossly loose and was revised without complications.The tibial and patellar components were well-fixed and remained intact.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: revision operative notes confirm that patient was revised due to loosening, pain, effusions and laxity.Femoral component was found to be loose and was extracted without difficulty.Total of nine pre-operative and post-operative images were reviewed and not submitted to mmi as there is sufficient documentation within the medical records and submission would not enhance the investigation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL POSTERIOR AUGMENT CEMENTED SIZE 11, 11+ 5 MM THICKNESS
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15988254
Report Number0001822565-2022-03502
Device Sequence Number5961871
Product Code JWH
UDI-Device Identifier00889024556942
UDI-Public(01)00889024556942(17)300914(10)64804890
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K191625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42556807005
Device Lot Number64804890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer03/06/2023
04/26/2024
10/29/2024
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date03/09/2023
05/09/2024
10/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Other;
Patient SexMale
Patient Weight83 KG
Patient RaceWhite
-
-