C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER Back to Search Results
Model Number 0170SI16
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that patient said this was not the correct catheter and these silicone coude foley catheters did not work, and they were hard for the patient to use.Patient¿s doctor had to use a wire to insert the catheters in place.The catheter that patient needed were on back order.
 
Event or Problem Description
It was reported that patient said this was not the correct catheter and these silicone coude foley catheters did not work, and they were hard for the patient to use.Patient¿s doctor had to use a wire to insert the catheters in place.The catheter that patient needed were on back order.
 
Additional Manufacturer Narrative
The reported event is unconfirmed as the product meets specifications.Visual evaluation noted received a 2-way all silicone catheter in unopened packaging with no obvious defects present.Inflated balloon with 10 ml of solution and the catheter rested with no leaks.Deflated balloon passively with no cuffing or leaks noted.Therefore, product meets specifications which states "missing/incorrect components are not allowed." the product was used for patient diagnostic or treatment.The product was not influenced by the reported event.A dhr review is not required as the reported event is unconfirmed.A labelling review is not required as the reported event is unconfirmed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15988542
Report Number1018233-2022-09559
Device Sequence Number5922892
Product Code EZL
UDI-Device Identifier00801741025341
UDI-Public(01)00801741025341
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number0170SI16
Device Catalogue Number0170SI16
Device Lot NumberNGGT3407
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer03/13/2023
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-