| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Eye Burn (2523)
|
| Date of Event |
11/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: a field service engineer (fse) inspected the handpiece and confirmed that handpiece had no issue.All results were within specifications.Manufacturing record review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that thermal burn occurred during cataract surgery on a patients unknown eye.The procedure was successfully completed with backup, corneal incision 2 4mm.No surgical intervention was required.There was no patient injury.The patient was prescribed rinderon and diamox.It was stated that the patient is recovering postoperatively.Intraocular pressure had temporarily been slightly high, but it is decreasing.
|
| |
|