ETHICON INC. VCL CT BRD VIO 27IN 1 S/A CT-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J341H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a robotic total hysterectomy procedure on (b)(6) 2022 and barbed suture was used.During the procedure, when trying to get the needle out of the trocar, it got caught and the needle got stuck in the trocar.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The following information was requested, but unavailable: "* could you please clarify if the surgery was successfully completed? please provide more details.No further information is available.What is the lot number? no further information is available.Device evaluated by mfr: analysis: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that it was received one needle-suture piece that pertain to product code j341h.Upon initial inspection, it was observed that the needle-suture piece was received inside a trocar.The needle suture was removed and the swage and attachment area was noted to be as expected and no issues related to pull off were observed.The suture was examined, and the end of the suture was noted to be cut probably caused by a surgical instrument.In addition, body fluids were noted along the strand.In addition, ten barbs were measured, and all barbs met the requirements.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 27IN 1 S/A CT-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15988843
Report Number2210968-2022-10315
Device Sequence Number12314324
Product Code GAM
UDI-Device Identifier10705031039056
UDI-Public10705031039056
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberJ341H
Device Catalogue NumberJ341H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Initial Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-