EXACTECH, INC. GXL NUETRAL LINER, G0 28MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
GXL NUETRAL LINER, G0 28MM ID |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Joint Dislocation (2374)
|
| Date of Event |
09/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported by legal brief, on or about (b)(6) 2018, patient underwent a right hip arthroplasty due to severe osteoarthritis in her right hip.Orthopedic surgeon utilized exactech products in patient right hip and did not breach any generally accepted standard of care in the field of orthopedic surgery in his care and treatment of patient negligently cause any injury to her.During the surgery on or about november 28, 2018, patient had an exactech novation crown cup system implanted in her right hip, which included the following exactech components (hereinafter "the device"): (1) 48 mm novation crown cup; (2) 28 mm novation gxl liner; (3) 28 mm biolox delta femoral head; and (4) size 1 alteon tapered wedge femoral stem.At the time of implant, each of the exactech components of patient's right hip was in substantially the same condition in all relevant respects as when they left defendants' control.On or about (b)(6) 2022, patient underwent right hip revision surgery due to right total hip dislocation and polyethylene wear.Surgeon noted in the revision operative report that "with the hip reduced on fluoro with a non-weightbearing film, you could still appreciate a slight eccentricity of the femoral head within the acetabular liner and cup, and this is suspicious for polyethylene wear".Patient continues to endure painful recovery and rehabilitation process from her revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant products: biolox delta femoral head 28mm od, -3.5mm (cat# 170-28-93 / serial# (b)(4)).Cup, cluster-hole, 46mm group 0 (cat# 180-01-46 / serial# (b)(4)).Wedge plasma x/o sz 1 (cat# 188-01-01 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation results - based on the available information, the patient meets the following risk criteria for early prosthesis wear/osteolysis: significant edge loading of the femoral head on the acetabular liner (dislocation) and combination of largest available femoral head and/or thinnest available acetabular liner was used.The most likely cause for the revision reported due to early prosthesis wear is a combination of risk factors.However, prosthesis wear cannot be confirmed as radiographs were not provided and the devices were not available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b2, h6 - health effect - clinical code, medical device problem code, component code, h7 and h11 the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Based on the available information, the patient meets the following risk criteria for early prosthesis wear/osteolysis: significant edge loading of the femoral head on the acetabular liner (dislocation) and combination of largest available femoral head and/or thinnest available acetabular liner was used.The most likely cause for the revision reported due to early prosthesis wear is a combination of risk factors.However, prosthesis wear cannot be confirmed.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|