Model Number N/A
Medical Device Problem Codes Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Skin Inflammation/ Irritation (4545)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the patient that the cover patches are irritating her skin as well.The patient stated that she has sensitive skin, and she contacted her doctor who told her to use paper tape, which also irritated her skin.The patient also requested more electrodes.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Date of event: the event occurred sometime in (b)(6) 2022.Medical product: unknown.Therapy date: unknown.Customer has indicated that the product is in process of being returned to zimvie for investigation.Once the investigation is complete, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported by the patient that the cover patches are irritating her skin as well.The patient stated that she has sensitive skin, and she contacted her doctor who told her to use paper tape, which also irritated her skin.The patient also requested more electrodes.No additional patient consequences have been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The device not was returned to zimvie for investigation.The reported event was not verifiable after the investigation associated with skin irritation.The device history record was reviewed and no discrepancies related to the reported event were found.No physical and/or functional condition could be found after the dhr that could be considered a causal factor for the reported complaint.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly zimvie will continue to monitor for trends.B4: date of this report added.G1: contact office updated.G3: date received by manufacturer added.G6: type of report.H2: follow up type.H3: device evaluated by manufacturer updated to no.H4: device manufacturer date added.H6: component codes added 772 - cover.H6: impact code added 4648 - insufficient information.H6: device code updated to 2682 - patient-device incompatibility.H6: investigation code added to 3331 - analysis of production records.H6: investigation code added to 4114 - device not returned.H6: investigation code added to 4119 ¿ insufficient information available.H6: investigation findings code added to 3221: no findings available.H10: additional narratives/data.The following sections have been corrected: h6: device code updated to 2993: adverse event without identified device or use problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODE COVER PATCHES
Common Device Name
COVER PATCHES
Manufacturer (Section D)
EBI, LLC.
1 gatehall dr
parsippany NJ 07054
Manufacturer (Section G)
EBI, LLC.
1 gatehall dr
parsippany NJ 07054
Manufacturer Contact
stephanie smith
1 gatehall dr
parsippany, NJ 07054
9732999300
MDR Report Key15988954
Report Number0002242816-2022-00156
Device Sequence Number5724251
Product Code LOF
UDI-Device Identifier00812301020195
UDI-Public00812301020195
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P850022/S017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number106130-17
Device Lot NumberN/A
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date03/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-