ONKOS SURGICAL ELEOS; COLLAR LOCKING RING Back to Search Results
Model Number HR-30001-03M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2022-00370, #3013450937-2022-00371, #3013450937-2022-00372, #3013450937-2022-00373, #3013450937-2022-00374, #3013450937-2022-00375, #3013450937-2022-00376, #3013450937-2022-00377, #3013450937-2022-00378, #3013450937-2022-00380, #3013450937-2022-00381.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: distal femur, two male-female midsections, tibial hinge component, poly spacer, and distal femur axial pin.The following implants remain implanted from the patient's original surgery: biogrip collar, locking ring, segmental stem, tibial baseplate, canal-filling stem extension, and resurfacing patella.No additional information regarding this adverse event has been reported.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: distal femur, two male-female midsections, tibial hinge component, poly spacer, and distal femur axial pin.The following implants remain implanted from the patient's original surgery: biogrip collar, locking ring, segmental stem, tibial baseplate, canal-filling stem extension, and resurfacing patella.No additional information regarding this adverse event has been reported.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
COLLAR LOCKING RING
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
PHILIPS PRECISION INC
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key15990459
Report Number3013450937-2022-00379
Device Sequence Number6103018
Product Code KRO
UDI-Device IdentifierB278HR3000103M0
UDI-PublicB278HR3000103M0
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K203815
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHR-30001-03M
Device Catalogue NumberHR-30001-03M
Device Lot Number88776R-073
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000009E, ELEOS DISTAL FEMUR; P/N 25001040E, ELEOS MALE FEMALE MIDSECTION; P/N 25001040E, ELEOS MALE FEMALE MIDSECTION; P/N 25001208E, ELEOS POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N HC-11120-03M, ELEOS SEGMENTAL STEM; P/N KPONTP35E, ELEOS RESURFACING PATELLA; P/N KSP12100E, ELEOS CANAL FILLING STEM EXTENSION; P/N PB-3200R-03M, ELEOS BIOGRIP COLLAR; P/N TB-2203E-01M, ELEOS TIBIAL BASEPLATE; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
Patient SexMale
-
-