Model Number 6R45B
Medical Device Problem Code Unintended Ejection (1234)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the load 6r45b placed in the ats45 stapler and when introduced into the cavity, the load 6r45b staples fell open into the abdominal cavity before being triggered by the stapler.Staples were removed from the cavity.There was no harm to the patient.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 1/6/2023.Investigation summary: the product was returned to ethicon endo-surgery for evaluation, along with three photos.Visual inspection was conducted on the return reload.During the visual analysis of the photos, the following was observed: the first photo shows a tyvek from top view, code 6r45b lot: 758a49 (exp.Date: 2027-03- 31).The second photo shows a cartridge from the top view that appears to have some drivers up at the proximal end.The third photo shows a reload from the pan area and the sled can be seen partially advanced.Visual analysis of the returned sample determined that the 6r45b reload was received with no apparent damaged and partially fired 1/10 and with the reload lock out spring normal.No functional test could be performed due to the condition of the reload.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number 758a49, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELOADS FOR LINEAR CUTTERS
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15990667
Report Number3005075853-2022-08691
Device Sequence Number5995525
Product Code GDW
UDI-Device Identifier30705036000840
UDI-Public10705036000846
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6R45B
Device Catalogue Number6R45B
Device Lot Number758A49
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer12/17/2022
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-