Unit gets very hot and is uncomfortable to use as per cs rep ana maria lepiana, it was reported by the sales rep that the patient finds the bhs unit uncomfortable and the unit gets very hot.Sales rep requested to switch to an opak, he was advised to forward a new rx and dr not explaining why there should be a switch.(b)(6) received the signed and dated rx and note from the doctor., sales rep was advised he can apply the opak from his trunk stock and to pick up the complete bhs unit including the coils from the patient.A return label was sent to the sales rep.
|
|
Corrections in b4: date of the report, b5: event description, d3: manufacturer g1: contact office, g6: type of report, h2, h8 and h3.Additional information in d4, d8-9, g3, h4: device manufacture date, h6: component, investigation type, findings, and conclusions and h10: additional narrative.Investigation summary: the sflx 5 coil was returned for further evaluation.A visual inspection of the customer returned product was performed.Included was one sflx 5 coil part no.1068224 with (b)(6).The part appeared to be in good condition from the cosmetic/visual point of view.The failure was not confirmed for the reported condition of "unit gets very hot and is uncomfortable to use".The coil was tested and operates as intended.Review of complaint history identified (b)(4) total complaints from ((b)(6) 2021) to ((b)(6) 2022) for pn (1068224) and events related to (pain).Keyword search criteria: (complaint code: medical: pain) the search could not be specified further because the main complaint was pain.Review of the information provided by the customer and the findings from the investigation indicated that no physical and/or device functional condition could be found that could be considered a causal factor for the reported complaint of "pain".No failure and/or fault condition could be found and confirmed.Therefore, no further actions are required at this time.No problem was found with the bhs unit and sflx coil related to unit getting hot during the investigation.Based on the information provided by the patient and investigation results, no permanent damage or impairment was reported due to uncomfortable and unit getting hot.Therefore, the reportability of the event has been made non-reportable at this time.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|