| Model Number |
6199949 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation determined that discordant reactive vitros sars-cov-2 antigen (cv2ag) results were obtained from a single patient sample when tested using vitros cv2ag lot 2040 on a vitros xt 7600 integrated system.The results were considered discordant as a negative pcr result was obtained for the patient.Additionally, a non-reactive vitros cv2ag result was obtained when the sample from the patient was retested.A definitive cause of the discordant false reactive vitros cv2ag results was not determined.A vitros cv2ag lot 2040 reagent issue is not a likely contributor to the event as historical qc results as well as results on the date of the event were within acceptable guidelines.In addition, ongoing tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros cv2ag lot 2040.An instrument issue cannot be completely ruled out as a contributor to the event as diagnostic precision testing to verify instrument performance was not completed.However, there was no indication of an instrument related issue as an expected vitros cv2ag result was obtained for the patient through repeat testing without any known actions being performed on the vitros xt 7600 system.The initial nasopharyngeal swab sample obtained from the patient was described as cloudy and viscous by the customer.Therefore, a sample related issue cannot be completely ruled out as a contributor to the event.Additionally, the initial vitros reactive result was reproducible as repeat testing of the initial extract was also reactive.However, after centrifugation of the initial sample, a non-reactive result was obtained.Therefore, pre-analytical sample processing cannot be ruled out as a contributing factor of the event.
|
| |
|
Event or Problem Description
|
|
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report discordant reactive vitros sars-cov-2 antigen (cv2ag) results were obtained from a single patient sample when tested using vitros cv2ag lot 2040 on a vitros xt 7600 integrated system.The results were considered discordant as a negative pcr result was obtained for the patient.Additionally, a non-reactive vitros cv2ag result was obtained when the sample from the patient was retested.Vitros cv2ag results of reactive (1.31 and 1.07 s/c) versus expected result of non-reactive biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected.The false reactive vitros cv2ag results were reported from the laboratory to a physician.However, no treatment was initiated, altered or stopped based on the reported results.There was no allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4) and reportability assessment (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Ortho received an email from fda on 25 jan 2023.
|
| |
|
Event or Problem Description
|
|
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report discordant reactive vitros sars-cov-2 antigen (cv2ag) results were obtained from a single patient sample when tested using vitros cv2ag lot 2040 on a vitros xt 7600 integrated system.The results were considered discordant as a negative pcr result was obtained for the patient.Additionally, a non-reactive vitros cv2ag result was obtained when the sample from the patient was retested.Vitros cv2ag results of reactive (1.31 and 1.07 s/c) versus expected result of non-reactive biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected.The false reactive vitros cv2ag results were reported from the laboratory to a physician.However, no treatment was initiated, altered or stopped based on the reported results.There was no allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4) and reportability assessment 600315.
|
| |
|