• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BRAEMAR MANUFACTURING, LLC UNIT, C6M, A10E; UNIT, C6M, A10E U

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BRAEMAR MANUFACTURING, LLC UNIT, C6M, A10E; UNIT, C6M, A10E U Back to Search Results
Model Number 02-01804
Medical Device Problem Codes Thermal Decomposition of Device (1071); Electrical /Electronic Property Problem (1198); Temperature Problem (3022); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer alleged that they received a melted cord but continued to use it and it smelled like it was burning when she plugged it in.Patient was worried that the monitor was damaged.
 
Additional Manufacturer Narrative
Engineering evaluation was completed results are as follows: device was inspected for general physical integrity and the charging port was inspected for evidence of damage.No problems found.Device started charge at 11:22am at 1442.2ma, device was charged fully at 12:34pm at 356.4ma, max temperature at 82.5f, normal.Total device charging time: 72 minutes.Device charged fully; device showed no signs of defect that could have contributed to overheat during charging see attachment 1 for results.The temperature test was not performed as the overheat event happened during charging.Device was production tested; no problems found.The charging cord that was being used at the time of the event was not returned.Although it is more probable that the allegedly melted cord was damaged and experienced a fault in which allowed excess current to flow, resulting in an overheat.Conclusion: engineering evaluation was unable to replicate overheat event.No problems found.Patient was sent a new monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIT, C6M, A10E
Common Device Name
UNIT, C6M, A10E U
Manufacturer (Section D)
BRAEMAR MANUFACTURING, LLC
1285 corporate center drive
suite 150
eagan MN 55121
Manufacturer (Section G)
BRAEMAR MANUFACTURING, LLC
1285 corporate center dr
suite 150
eagan MN 56121
Manufacturer Contact
beverly okoh
1285 corporate center drive
suite 150
eagan, MN 55121
6124263781
MDR Report Key15992427
Report Number2133409-2022-00016
Device Sequence Number17694382
Product Code DSI
UDI-Device IdentifierB146C6M70
UDI-PublicB146C6M70
Combination Product (Y/N)N
PMA/510(K) Number
K171410
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number02-01804
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2022
Initial Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-