| Model Number |
G2200 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during preparation for a water vapor therapy procedure, that the generator encountered continuous prime error messages with two delivery devices.The procedure could not be completed.There were no patient complications.This event is being reported for aborted/cancelled procedure with a patient under anesthesia or sedation is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation for a water vapor therapy procedure, that the generator encountered continuous prime error messages with two delivery devices.The procedure could not be completed.There were no patient complications.This event is being reported for aborted/cancelled procedure with a patient under anesthesia or sedation is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to field g1: manufacturing site upon receipt of this generator at our quality assurance laboratory, this device was thoroughly analyzed.During functional evaluation of the device, no error codes appeared.The generator passed the post (power on self-test).Syringe and delivery device were connected with no issues.Generator primed and flush as per specifications.Error 242 (pressure value below 250 mmhg during vapor generation) was found during evaluation.The generator received all the required updates and passed full functional and electrical safety testing.Based on the information available and analysis results, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Search Alerts/Recalls
|