| Model Number |
RT134 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint rt134 adult breathing circuit is currently en route to fisher & paykel healthcare in new zealand for investigation.We will provide a follow up report upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in japan that an rt134 adult breathing circuit did not pass the ventilator leak test before use.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in japan that an rt134 adult breathing circuit did not pass the ventilator leak test before use.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint rt134 adult breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected and leak tested.Results: visual inspection of the complaint rt134 adult breathing circuit revealed no damage was found to any of the breathing circuits.The leakage test also revealed that the breathing circuit was within specification.Conclusion: we were unable to determine what may have caused the failed leak test as reported by the customer, as no fault was found with the returned device.All rt134 adult breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt134 adult breathing circuit state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
|
| |
|