Model Number RT134
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The complaint rt134 adult breathing circuit is currently en route to fisher & paykel healthcare in new zealand for investigation.We will provide a follow up report upon completion of investigation.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan that an rt134 adult breathing circuit did not pass the ventilator leak test before use.There was no patient involvement.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan that an rt134 adult breathing circuit did not pass the ventilator leak test before use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint rt134 adult breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected and leak tested.Results: visual inspection of the complaint rt134 adult breathing circuit revealed no damage was found to any of the breathing circuits.The leakage test also revealed that the breathing circuit was within specification.Conclusion: we were unable to determine what may have caused the failed leak test as reported by the customer, as no fault was found with the returned device.All rt134 adult breathing circuits are visually inspected, and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt134 adult breathing circuit state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON-HEATED ADULT BREATHING CIRCUIT
Common Device Name
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key15992875
Report Number9611451-2022-01199
Device Sequence Number14884003
Product Code BTT
UDI-Device Identifier09420012430533
UDI-Public(01)09420012430533(10)2101720405(11)210714
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131895
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT134
Device Catalogue NumberRT134
Device Lot Number2101720405
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2022
Supplement Date Received by Manufacturer04/18/2023
Initial Report FDA Received Date12/15/2022
Supplement Report FDA Received Date04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-