| Model Number |
80619 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the catheter split in half.A direxion infusion catheter was selected for use in an embolization procedure.The catheter was introduced into the patient.However, it was noted that the catheter appeared damaged.It was removed, and upon inspection, it was found kinked and split in half.A new direxion infusion catheter was used to complete the procedure.There were no patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter split in half.A direxion hi-flo infusion catheter was selected for use in an embolization procedure.The catheter was introduced into the patient.However, it was noted that the catheter appeared damaged.It was removed, and upon inspection, it was found kinked and split in half.A new direxion hi-flo was used to complete the procedure.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by mfr: the direxion microcatheter was returned for analysis.The device shaft was visually and microscopically analyzed for any damage.The devices shaft showed a fracture located 8.8cm from the hub.The device was not completely separated the inner liner was still attached.No kinks were noticed.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
| |
|