Model Number MMT-1712KL
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2022
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the customer received a pump error 130.This alarm is generated when the pump detects movement in hall sensor counts that are unexpected since the pump did not command motor movement.Troubleshooting was performed and the alarm was cleared successfully and the rewind completed.Customer stated that they performed displacement test and self test.Insulin pump passed it.No harm requiring medical intervention was reported.  it was unknown whether the customer continued using the device or not.Insulin pump will not be returned for analysis.
 
Manufacturer Narrative
(b)(4).The insulin pump involved in this event is the ngp 640g insulin pump which is not marketed in the united states.However, the device is similar to the ngp insulin pump, which is marketed in the united states.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1712KL 640G V4.10 BK SF MG
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key15993378
MDR Text Key308072839
Report Number2032227-2022-379752
Device Sequence Number1
Product Code OYC
UDI-Device Identifier000000763000317195
UDI-Public(01)000000763000317195
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712KL
Device Catalogue NumberMMT-1712KL
Device Lot NumberHG5QDAK
Was Device Available for Evaluation? No
Date Manufacturer Received11/25/2022
Date Device Manufactured09/21/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-