PERFUSION SYSTEMS CANNULA CB77522 EOPA BLUNT 22FR CARMEDA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
| Model Number |
CB77522 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that during use of a eopa cannula, after cannulating the patient, customer found that the cannula was leaking.The device was replaced to complete the procedure.There was no patient impact associated with this event.Medtronic received additional information that the patient lost approximately 50 ml of blood as a result of this leak.No blood tran sfusion was required.The cannula was not heated or cooled prior to use.The leak was from the cannula, from a hole between the wired part of the canula.The damage was not in the location of the sutures.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection shows evidence of damage/cut in the device.During the cleaning process there was a leak from the damaged location.Reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|