• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION TRAUMAONE SYSTEM SCREW 2.0X12MM CROSS-DRIVE; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TRAUMAONE SYSTEM SCREW 2.0X12MM CROSS-DRIVE; PLATE, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/19/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the placement of the screw for fixation of a mandibular fracture, the screw head fractured, and the screw body was retained in the bone.The head of the screw remained in the screwdriver.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Report source: colombia.Provided photos of screw appears to have fractured at the base of the head of the screw near the thread area.No other determination can be made by the supplied photos.No product was returned; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A visual inspection was conducted on the returned screw.The screw shows signs of attempted use.There is marking on the screw head.The screw shaft has fractured where it meets the screw head.The complaint is confirmed.A determination cannot be made as to what caused the screw to fracture.A definitive root cause cannot be determined.This report is being submitted to update additional information in section b4, b5, d9, g3, g6, h2, h3, and h10.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMAONE SYSTEM SCREW 2.0X12MM CROSS-DRIVE
Common Device Name
PLATE, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15995228
Report Number0001032347-2022-00367
Device Sequence Number14939886
Product Code JEY
UDI-Device Identifier00888233000284
UDI-Public(01)00888233000284(10)J722580
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K081067
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number41-2012
Device Lot NumberJ722580
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2021
Supplement Date Received by Manufacturer03/03/2023
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-