The device is returned, and an evaluation completed for it.The user¿s complaint was confirmed.Upon inspection and testing, it was observed that probe is broken off at 16.8 mm from the distal end site.The broken piece of the probe tip was not returned.Since the breakage is at the proximal end site, it is not possible to check if there any contact mark on inner site surface of the probe.The device h-electrode has coating peeled and scratched.The tissue pad has minor deformation.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
|
Additional information has been received for this event from the customer.There is more information on the device evaluation this supplemental report is being submitted to provide this information.Please see the updates in sections: b5, g3, g6, h2, h4, h6, and h10.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria at inspection.The breakage of the probe started at the cracked area of the probe.Also, black discoloration of the area near the starting point of the breakage was observed.No scratches or contact marks on the non-insulated area of the grasping section were observed.The exact cause of the breakage of the probe could not be determined.However, based on device evaluation and the results of investigation in the past, a likely mechanism causing the breakage of the probe might be the following: during the output activation in seal and cut mode, the distal end of the probe was slightly contacting a device with a thin tip, causing spark generation.A force was applied to the probe during spark generation.A force to grasp tissue or a force to activate the output in seal and cut mode was applied to the probe.This generated cracks on the probe.Due to continuous use of the device, the cracks on the probe progressed.This led to breakage of the probe.The instructions for use includes the following statements that warn against the issue: do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.
|