After reviewing the system data logs, the rapidpoint instrument in question was found to be functioning as intended.There were no system or sensor related error events in and around the time of the escalated samples.The qc performance was found to be stable throughout the use life of the installed cartridge.For this patient, there were two extreme k+ recoveries on (b)(6).However, the rapidpoint and lab instrument potassium agreed for the samples analyzed within 10 minutes of each other.The cause of this event is unknown.
|