U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE II XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE II XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S200-IR
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was not confirmed.The evaluation uncovered a faulty emergency lamp, faulty hinge, retaining bracket 180, slide detection 180 and spring 180, internal condition was found to be ok, rear foot was found corroded, corrosion on rear panel, corrosion and paint mark on top cover, corrosion on link-out connector, but working ok, minor corrosion on chassis and minor scratch and hit marks on the front panel.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported to olympus that during maintenance, a e207(emergency lamp is blown or failed) and spare lamp indication glowing on the visera elite ii xenon light source.There was no patient involvement associated with this event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation and to correctly indicate that the customer's allegation was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, error e207 for emergency lamp occurred due to an emergency light failure.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: troubleshooting from error codes: coding: e207 error message: clv emergency light failure e207 contact olympus phenomenon: error that occurs when the emergency light of the light source unit is turned off.Check 1: emergency lamp operation error¿.Tier 2 =(1)e207 failure due to an emergency light failure (2)the emergency light indicator on the front panel blinks due to an emergency light failure." however, the definitive root cause of the emergency light failure could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE II XENON LIGHT SOURCE
Common Device Name
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15995796
Report Number3002808148-2022-05302
Device Sequence Number5961606
Product Code OWN
UDI-Device Identifier04953170428142
UDI-Public04953170428142
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K200542
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCLV-S200-IR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer12/20/2022
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/06/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-