• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS SCHON XL DIALYSIS CATHETER; SCHON¿ HEMODIALYSIS DIALYSIS CATHETER SETS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS SCHON XL DIALYSIS CATHETER; SCHON¿ HEMODIALYSIS DIALYSIS CATHETER SETS Back to Search Results
Catalog Number 10801701
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
An end user experienced an issue when using a schon xl 15cm.It was reported that when a patient went to dialysis for treatment, there were holes noted in the dialysis catheter.The catheter is presumed to have been removed and replaced in order for the patient to continue dialysis treatment.
 
Additional Manufacturer Narrative
Returned for evaluation was one (1) used schon dialysis catheter.As reported, a hole in extension tubing was confirmed in the blue luer lumen extension tube.The customer's reported complaint of a "hole in the blue port catheter/extension leg" was confirmed.The returned sample was placed in water and air was applied to the catheter and the hole was confirmed in the extension leg.Scar004679 and the dialysis catheter complaint sample was sent to supplier/manufacturer (martech) for sample evaluation/root cause determination and dhr review of the reported lot number.Per scar004679 response, the hole in the extension tubing is consistent with those generated by the clamp when closing it with the guidewire still inside the catheter [extension leg tubing].Dhr review showed no issues of concern.The root cause of the hole in the extension leg tubing is likely due to handling damage by end user closing device clamp with guidewire inside of the extension leg tubing.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.  the review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use, which is supplied to the end user with this catalog number, states: "clamping the catheter repeatedly in the same spot could weaken the tubing.Change the position of the clamps regularly to prolong the life of the tubing." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference: (b)(4).
 
Additional Manufacturer Narrative
Correction: d4.Additional information: device received.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCHON XL DIALYSIS CATHETER
Common Device Name
SCHON¿ HEMODIALYSIS DIALYSIS CATHETER SETS
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15996030
Report Number1319211-2022-00056
Device Sequence Number5995748
Product Code MSD
UDI-Device IdentifierH787108017015
UDI-PublicH787108017015
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K971925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2026
Device Catalogue Number10801701
Device Lot NumberMPWG120
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2022
Supplement Date Received by Manufacturer12/29/2022
03/09/2023
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date01/24/2023
03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-